Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: A.W. Sperry DSA-440T: (1 year ) CAL VER /5520 DATE: 11-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 12 NUMBER OF LINES: 88 CONFIGURATION: Fluke 5520A STANDARD: Fluke 5500A/COIL ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F V 1.002 ASK+ K X 1.003 STD Fluke 5500A/COIL 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: NONE 1.005 DISP Temperature: 25 +/- 5degC 1.005 DISP Humidity: <75% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD {DC VOLTAGE} 1.008 DISP Connect the 5520A to the UUT as follows: 1.008 DISP [32] 5520A NORMAL HI to UUT V/OHM 1.008 DISP [32] 5520A NORMAL LO to UUT COM 1.009 HEAD DC VOLTAGE: {4V Range} 1.010 5520 4 3.500V 0.5% 0.001U 2W 2.001 HEAD DC VOLTAGE: {400V Range} 2.002 5520 400 350.0V 0.5% 0.1U 2W 3.001 HEAD {AC VOLTAGE} 3.002 HEAD AC VOLTAGE: {400V Range} 3.003 5520 400 350.0V 1.2% 0.4U 60H SI 2W 4.001 HEAD AC VOLTAGE: {750V Range} 4.002 5520 750 700V 1.5% 4U 60H SI 2W 5.001 HEAD {DIODE} 5.002 DISP Set the FUNCTION/RANGE switch to DIODE. 5.003 5520 4 0.500V 0.1U 2W 6.001 ASK+ R 6.002 DISP Connect the 5520A to the UUT as follows: 6.002 DISP [32] 5520A AUX HI to UUT V/OHM 6.002 DISP [32] 5520A AUX LO to UUT COM 6.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 6.002 DISP [32] 5520A NORMAL LO to UUT COM 6.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 6.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 6.003 5520 10Z S CW 6.004 EVAL Is an audible tone present {(Continuity test OK)}? 7.001 HEAD {RESISTANCE} 7.002 HEAD RESISTANCE: {400 Ohm Range} 7.003 5520 400 300.0Z 1% 0.1U CW 8.001 HEAD {AC CURRENT} 8.002 HEAD AC CURRENT: {40A Range} 8.003 DISP Connect the 5500A/COIL to the 5520A AUX HI and LO. 8.003 DISP Clamp the UUT around the 5500A/COIL. 8.004 5520 40 30.00A 2% 0.04U 60H SI T5 9.001 HEAD AC CURRENT: {400A Range} 9.002 DISP Connect the 5500A/COIL to the 5520A 20A and AUX LO. 9.002 DISP Clamp the UUT around the 5500A/COIL. 9.003 5520 400 300A 2% 4U 60H SI HC T5 10.001 HEAD {TEMPERATURE} 10.002 DISP Set the FUNCTION/RANGE switch to TEMP (degC). 10.003 DISP Connect a K-type thermocouple to the TEMP jack 10.003 DISP on the DSA-440T. Connect the other end of the 10.003 DISP thermocouple to the TC terminals on the 5520A. 10.004 M5520 TCREF 21degC 10.005 5520 -20degC 5% 3U _K TC 2W 11.001 5520 700degC 3% 1U _K TC 2W 12.001 END